The Nuvair 9507 replacement oxygen sensor is for use with the DE-OX ZIP PRO oxygen analyzer. Optional diverter available for purchase.
The Nuvair 9507 is a replacement electrochemical oxygen sensor for gas analyzers. Under normal use, oxygen sensors have an expected service life of 36 months. The 9507 replacement sensor fits the DE-OX ZIP PRO gas analyzer.
Optional diverter available for purchase (SKU DIVERTER, now black in color).
WARNING: Never expose gas sensors to pressure or you may cause damage and/or false readings. Damaged sensors will not provide accurate gas analysis. Most gas analyzers can be used to analyze a regulated gas sample flow, the contents of a gas cylinder, or the flow from a regulator. The flow rate of gas must equal 1-5 L/min. To produce this flow, a Flow Restrictor and Regulator may be required. A faulty Flow Restrictor can lead to a false analyzer reading. Flow Restrictors should be regularly tested with a Flow Meter. Inaccurate gas analysis can lead to serious personal injury or death.
Output | 8-13 mV in air at 25°C/77°F, sea level |
---|---|
Electrical Connector | Typical mating plug switchcraft: .141" miniature phone plugs - screw or clamp-lug terminal Polarity: Tip = negative (–), Sleeve = positive (+) |
Range | 0-100% Oxygen |
Accuracy | ±1% of full scale at constant temperature and pressure (0-1 atm) when calibrated with 100% oxygen |
Sensor Type | Electrochemical |
External Thread Size | M 16X1 |
Expected Sensor Life | 36 months in air at 25°C/77°F and 50% rh |
Response Time | <6 seconds for 90% of final value |
Operating Temperature | 32-104°F (0-40°C) |
Storage Temperature | 32-122°F (0-50°C) |
Humidity | 0-99% rh (non-condensing) |
Weight | 1.1 oz (31 g) |
Application | For surface/above-water diving cylinder verification applications; not intended for underwater use. |
Warranty | 0 – 12 Months Free Replacement 13 – 18 Months 50% Off Replacement 19 – 24 Months 25% Off Replacement |